Type | Benchtop fume hoods |
Exhaust system | CAV exhaust system (Constant air volume) |
Exhaust fan | Inclined flow fan (The installation principle is the same as air conditioning) |
Exhaust air rate | 1500-3500m3/h |
Ventilation ducts | PVC material, FPR, PP or other |
Power | (Single-phase) 220V/50HZ or (Three-phase) 380V/50HZ |
Air valve | Manual or electric |
Average wind speed | 0.3-0.6m/s adjustalbe |
Noise level | < 60dB(A) |
Tops | 12.7/16/19/mm thickness of Phenolic resin (Other: Epoxy resin, Ceramic etc) |
Inner Liner | Using HPL physiochemical board to make for lining plate and the flow guide plate |
Cabinet | 1.0mm thick cold-rolled steel with 75μm epoxy resin powder coating, most chemical resistant |
Size | 1200/1500/1800*850*2350 mm |
Visual windows | Transparent tempered glass, explosion-proof (manual) |
Opening height | 20-700 mm |
Control panel | Touch, intelligent control panel (can control lights, sockets, exhaust fan etc) |
Gross weight | 280-400 kg |
Motor power | 0.5-1.1KW |
Luminance | >300LX |
Lab accessories | Pegboard, sink, faucet, gas fitting, sockets, lab stools etc |